MCA527
Command name CMD_SET_PUR
Compatibility Identical to the MCA166 command.
Execution right Necessary
Description The command sets the pile up rejection.
pur = 0 turns PUR off
pur 0 turns PUR on
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer long integer
Parameter preamble command pur 0 end flag
Byte string (HEX) A5 5A 53 00 pur 00 00 00 00 00 B9 9B
Command name
CMD_SET_FAST
Compatibility
Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software
Execution right Necessary
Description This command sets the fast discriminator threshold (0 ... 2499, default 400).
Format integer integer integer long integer
Parameter Preamble command fast 0 end flag
Byte string (HEX) A5 5A 50 00 fast
l
fast
h
00 00 00 00 B9 9B
Command name
CMD_SET_SLOW
Compatibility
Compatible to the MCA166 command. The command is meaningless for the MCA527. It has only
been retained for compatibility with older software
Execution right Necessary
Description This command sets the slow discriminator threshold (0 ... 2499, default 400).
Format integer integer integer long integer
Parameter Preamble command slow 0 end flag
Byte string (HEX) A5 5A 51 00 slow
l
slow
h
00 00 00 00 B9 9B