Gbs-elektronik MCA-527 Firmware Commands Manual do Utilizador Página 27

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 54
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 26
Description of the MCA527 Firmware Commands
Command name CMD_STOP_EXTENSION_PULSER
Compatibility New MCA527 command.
Execution right Necessary
Description The command stops the selected pulser. The command will only be executed successfully if the
corresponding parts have been correctly configured (see CMD_SET_EXTENSION_PORT).
part: 1 = part B (pulser2)
3 = part D (pulser1)
7 = part B (pulser2) and part D (pulser1)
Format integer integer integer long integer
Parameter preamble command part 0 end flag
Byte string (HEX) A5 5A 23 01 part 00 00 00 00 00 B9 9B
Command name
CMD_SET_EXTENSION_OUTPUT
Compatibility New MCA527 command.
Execution right Necessary
Description The command sets the selected output. The command will only be executed successfully if the
corresponding parts have been correctly configured (see CMD_SET_EXTENSION_PORT).
part: 1 = part B (output2)
3 = part D (output1)
7 = part B (output2) and part D (output1)
o1: 0 = off, ≠0 = on
o2: 0 = off, ≠0 = on
Format integer integer integer integer integer integer
Parameter preamble command part o1 o2 end flag
Byte string (HEX) A5 5A 24 01 part 00 o1 00 o2 00 B9 9B
MCA File System Command:
Command name
CMD_WRITE_FILE
Compatibility New MCA527 command.
Execution right Necessary
Description The command writes the current measurement into a file. The file name is created automatically from
a consecutive number. Basis for this command is that a microSD card with enough free memory is
plugged in.
The command is ignored and responds with an error value if a measurement is still running.
Format integer integer integer integer integer integer
Parameter preamble command 0 0 0 end flag
Byte string (HEX) A5 5A 28 01 00 00 00 00 00 00 B9 9B
Remarks Call CMD_QUERY_STATE527_EX to get the needed and the free microSD memory size. The
command returns immediately., so CMD_QUERY_STATE527_EX has to be called again to get the
state and the result of the file writing process.
27
Vista de página 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 53 54

Comentários a estes Manuais

Sem comentários