Gbs-elektronik MCA-527 Firmware Commands Manual do Utilizador Página 34

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 54
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 33
MCA527
Command name CMD_QUERY_STATE (2
nd
Continuation)
Result data array
Byte offset 100 Start time unsigned long See CMD_START
Byte offset 104 TDF unsigned short always 0 at MCA527
Byte offset 106 Command flag and parameters 8 bytes
Byte offset 114 Read out buffer state unsigned short OCCUPIED = 0x2000
OVERRUN = 0x4000
FILLED = 0x8000
Byte offset 116 Counts per second
35
unsigned long
Byte offset 120 Differential dead time [ ‰ ] unsigned short
Byte offset 122 HV inhibit mode short
Byte offset 124 HV inhibit state unsigned short
Byte offset 126 Checksum unsigned short
Byte offset 128 MCA state unsigned short See CMD_QUERY_POWER
Byte offset 130 Start flag
36
unsigned short See CMD_START
35 Since firmware version 13.00.
36 Since firmware version 12.02.
34
Vista de página 33